Published in

Oxford University Press, EP Europace, 4(26), 2024

DOI: 10.1093/europace/euae087

Links

Tools

Export citation

Search in Google Scholar

Personalized pulmonary vein isolation with very high-power short-duration lesions guided by left atrial wall thickness: the QDOT-by-LAWT randomized trial

This paper was not found in any repository, but could be made available legally by the author.
This paper was not found in any repository, but could be made available legally by the author.

Full text: Unavailable

Green circle
Preprint: archiving allowed
Orange circle
Postprint: archiving restricted
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

Abstract Aims Pulmonary vein isolation (PVI) for paroxysmal atrial fibrillation (PAF) using very high-power short-duration (vHPSD) radiofrequency (RF) ablation proved to be safe and effective. However, vHPSD applications result in shallower lesions that might not be always transmural. Multidetector computed tomography-derived left atrial wall thickness (LAWT) maps could enable a thickness-guided switching from vHPSD to the standard-power ablation mode. The aim of this randomized trial was to compare the safety, the efficacy, and the efficiency of a LAWT-guided vHPSD PVI approach with those of the CLOSE protocol for PAF ablation (NCT04298177). Methods and results Consecutive patients referred for first-time PAF ablation were randomized on a 1:1 basis. In the QDOT-by-LAWT arm, for LAWT ≤2.5 mm, vHPSD ablation was performed; for points with LAWT > 2.5 mm, standard-power RF ablation titrating ablation index (AI) according to the local LAWT was performed. In the CLOSE arm, LAWT information was not available to the operator; ablation was performed according to the CLOSE study settings: AI ≥400 at the posterior wall and ≥550 at the anterior wall. A total of 162 patients were included. In the QDOT-by-LAWT group, a significant reduction in procedure time (40 vs. 70 min; P < 0.001) and RF time (6.6 vs. 25.7 min; P < 0.001) was observed. No difference was observed between the groups regarding complication rate (P = 0.99) and first-pass isolation (P = 0.99). At 12-month follow-up, no significant differences occurred in atrial arrhythmia-free survival between groups (P = 0.88). Conclusion LAWT-guided PVI combining vHPSD and standard-power ablation is not inferior to the CLOSE protocol in terms of 1-year atrial arrhythmia-free survival and demonstrated a reduction in procedural and RF times.