Nature Research, Scientific Reports, 1(11), 2021
DOI: 10.1038/s41598-021-01601-y
Full text: Download
AbstractHigh-sensitivity detection of hydrogen (H) contained in zircaloy-4, a commonly used material for nuclear fuel containers, is crucial in a nuclear power plant. Currently, H detection is performed via gas chromatography, which is an offline and destructive method. In this study, we developed a technique based on metastable excited-state He-assisted excitation to achieve excellent quality of H emission spectra in double-pulse orthogonal laser-induced breakdown spectroscopy (LIBS). The production of metastable excited-state He atoms is optimized by using LiF as sub-target material. The results show a narrow full-width-at-half-maximum of 0.5 Å for the H I 656.2 nm emission line, with a detection limit as low as 0.51 mg/kg. Thus, using this novel online method, H in zircaloy-4 can be detected efficiently, even at very low concentrations.