Full text: Unavailable
Abstract Objectives Non-invasive prenatal testing requires the presence of fetal DNA in maternal plasma. Understanding how preexamination conditions affect the integrity of cell-free DNA (cfDNA) and fetal fraction (FF) are a prerequisite for test implementation. Therefore, we examined the adjusted effect that EDTA and Streck tubes have on the cfDNA quantity and FF. Methods A total of 3,568 maternal blood samples across Canada were collected in either EDTA, or Streck tubes, and processing metrics, maternal body mass index (BMI), gestational age and fetal karyotype and sex were recorded. Plasma samples were sequenced using two different sequencing platforms in separate laboratories. Sequencing data were processed with SeqFF to estimate FF. Linear regression and multivariate imputation by chained equations were used to estimate the adjusted effect of tube type on cfDNA and FF. Results We found a positive association between cfDNA quantity and blood shipment time in EDTA tubes, which is significantly reduced with the use of Streck tubes. Furthermore, we show the storage of plasma at −80 °C is associated with a 4.4% annual relative decrease in cfDNA levels. FF was not associated with collection tube type when controlling for confounding variables. However, FF was positively associated with gestational age and trisomy 21, while negatively associated with BMI, male fetus, trisomy 18, Turners syndrome and triploidy. Conclusions Preexamination, maternal and fetal variables are associated with cfDNA quantity and FF. The consideration of these variables in future studies may help to reduce the number of pregnant women with inconclusive tests as a result of low FF.