Published in

MDPI, Biosensors, 4(12), p. 204, 2022

DOI: 10.3390/bios12040204

Links

Tools

Export citation

Search in Google Scholar

Label-Free Cholesteric Liquid Crystal Biosensing Chips for Heme Oxygenase-1 Detection within Cerebrospinal Fluid as an Effective Outcome Indicator for Spontaneous Subarachnoid Hemorrhage

Journal article published in 2022 by Hui-Tzung Luh ORCID, Yi-Wei Chung, Po-Yi Cho, Yu-Cheng Hsiao ORCID
This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Green circle
Published version: archiving allowed
Data provided by SHERPA/RoMEO

Abstract

A novel device for cholesteric liquid crystal (LC; CLC)-based biosensing chips for detecting heme oxygenase (HO)-1 within the cerebrospinal fluid (CSF) was invented. In the CLC device, the reorientation of the LCs was strongly influenced by the alignment layer surface and adjacent LCs. When the substrate was coated with the alignment layer, the CLCs oriented homeotropically in a focal conic state. Once HO-1 was immobilized onto the orientation sheet-coated substrate, the CLC changed from a focal conic state to a bright planar state by disrupting the CLCs. The concentration of HO-1 within CSF was shown to be an effective outcome indicator for patients with a spontaneous subarachnoid hemorrhage. We showed that the CLC immunoassaying can be used to measure HO-1 with a lower detection limit of about 10 ng/mL. The linear range was 10 ng/mL to 1 mg/mL. An easy-to-use, rapid-detection, and label-free CLC immunoassay device is proposed.