Experimental characterisation of the magnetic field correction factor, kB⃗, for Roos chambers in a parallel MRI-linac

Journal article published in 2022 by Jarrad Begg ORCID, Urszula Jelen ORCID, Paul Keall ORCID, Gary Liney, Lois Holloway ORCID
This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Green circle
Preprint: archiving allowed
Orange circle
Postprint: archiving restricted
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

Abstract Objective. Reference dosimetry on an MRI-linac requires a chamber specific magnetic field correction factor, k B ⃗ . This work aims to measure the correction factor for a parallel plate chamber on a parallel MRI-linac. Approach. k B ⃗ is defined as the ratio of the absorbed dose to water calibration coefficient in the presence of the magnetic field, N D , w B ⃗ relative to that under 0 T conditions, N D , w 0 T . k B ⃗ was measured via a N D , w transfer to a field chamber at each magnetic field strength from a chamber with known N D , w and k B ⃗ . This was achieved on the parallel MRI-linac by moving the measurement set-up between a high magnetic field strength region at the MRI-isocentre and a low magnetic field strength region at the end of the bore whilst maintaining consistent set-up and scatter conditions. Three PTW 34001 Roos chambers were investigated as well as a PTW 30013 Farmer used to validate methodology. Main Results. The beam quality used for the measurements of k B ⃗ was TPR 20/10 = 0.632. The k B ⃗ for the PTW Farmer chamber at 1 T on a parallel MRI-linac was 0.993 ± 0.013 (k = 1). The average k B ⃗ factor measured for the three Roos chambers on a 1 T parallel MRI-linac was 0.999 ± 0.014 (k = 1). Significance. The results presented are the first measurements of k B ⃗ for a Roos chamber on a parallel MRI-linac. The Roos chamber results demonstrate the potential for the chamber as a reference dosimeter in parallel MRI-linacs.