Published in

American Institute of Physics, Review of Scientific Instruments, 11(93), p. 113508, 2022

DOI: 10.1063/5.0100970

Links

Tools

Export citation

Search in Google Scholar

Ultra-compact x-ray spectrometer for high-repetition-rate laser–plasma experiments

This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Orange circle
Published version: archiving restricted
Data provided by SHERPA/RoMEO

Abstract

We present in this work the development of an ultra-compact, multi-channel x-ray spectrometer (UCXS). This diagnostic has been specially built and adapted to perform at high-repetition-rate (>1 Hz) for high-intensity, short-pulse laser plasma experiments. X-ray filters of varying materials and thicknesses are chosen to provide spectral resolution up to Δ E ≈ 1 keV over the x-ray energy range of 1–30 keV. These filters are distributed over a total of 25 channels, where each x-ray filter is coupled to a single scintillator. The UCXS is designed to detect and resolve a large variety of laser-driven x-ray sources such as low energy bremsstrahlung emission, fluorescence, and betatron radiation (up to 30 keV). Preliminary results from commissioning experiments at the ABL laser facility at Colorado State University are provided.