Autoantibodies Associated with Autoimmune Liver Diseases in a Healthy Population: Evaluation of a Commercial Immunoblot Test

This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Green circle
Published version: archiving allowed
Data provided by SHERPA/RoMEO

Abstract

Autoantibodies constitute important tools for diagnosing the autoimmune liver diseases (AILD) autoimmune hepatitis and primary biliary cholangitis. The EUROLINE immunoblot assay, detecting multiple specificities, is widely used, but the clinical importance of weakly positive findings is unclear. The manufacturer’s recommended cut-off was evaluated by investigating AILD-associated autoantibodies in 825 blood donors and 60 confirmed AILD cases. Positive findings were followed up with immunofluorescence microscopy on rat tissue, anti-M2-ELISA, alternative immunoblot assay, and liver function tests. Thirty-six (4.4%) blood donors were positive with EUROLINE. The most common specificities were LC-1 (1.6%), gp210 (1.3%), and AMA-M2 (1.1%). In general, the positive results were higher in patients than in blood donors, whereas anti-LC-1 was higher in blood donors. The liver function tests were slightly elevated in 2 of the 36 immunoblot positive blood donors. The majority of the positive EUROLINE findings could not be confirmed with the follow-up tests. The EUROLINE-Autoimmune Liver Diseases-(IgG) immunoblot detected autoantibodies in 4.4% of blood donors without signs of AILD. Our findings indicate that the recommended cut-off can be raised for most specificities without loss of diagnostic sensitivity. The prevalence of anti-LC-1 among blood donors indicates a problem with the antigen source.