SAGE Publications, Surgical Innovation, 3(29), p. 438-445, 2021
DOI: 10.1177/15533506211031459
Full text: Unavailable
Background: For centuries, surgeons have relied on surgical drains during postoperative care. Despite all advances in modern medicine and the area of digitalization, as of today, most if not all assessment of abdominal secretions excreted via surgical drains are carried out manually. We here introduce a novel integrated Smart Sensor System ( Smart Drain) that allows for real-time characterization and digitalization of postoperative abdominal drain output at the patient’s bedside. Methods: A prototype of the Smart Drain was developed using a sophisticated spectrometer for assessment of drain output. The prototype measures 10 × 6 × 6 cm and therefore easily fits at the bedside. At the time of measurement with our Smart Drain, the drain output was additionally sent off to be analyzed in our routine laboratory for typical markers of interest in abdominal surgery such as bilirubin, lipase, amylase, triglycerides, urea, protein, and red blood cells. A total of 45 samples from 19 patients were included. Results: The measurements generated were found to correlate with conventional laboratory measurements for bilirubin (r = .658, P = .000), lipase (r = .490, P = .002), amylase (r = .571, P = .000), triglycerides (r = .803, P = .000), urea (r = .326, P = .033), protein (r = .387, P = .012), and red blood cells (r = .904, P = .000). Conclusions: To our best knowledge, for the first time we describe a device using a sophisticated spectrometer that allows for real-time characterization and digitalization of postoperative abdominal drain output at the patient’s bedside.