Springer, Magnetic Resonance Materials in Physics, Biology and Medicine, 6(35), p. 943-951, 2022
DOI: 10.1007/s10334-022-01014-6
Full text: Unavailable
Abstract Object Improve shimming capabilities of ultra-high field systems, with addition of an accessible low-complexity B0 shim array for head MRI at 7 T. Materials and methods An eight channel B0 shim coil array was designed as a tradeoff between shimming improvement and construction complexity, to provide an easy to use shim array that can be employed with the standard 7 T head coil. The array was interfaced using an open-source eight-channel shim amplifier rack. Improvements in field homogeneity for whole-brain and slice-based shimming were compared to standard second-order shimming, and to more complex higher order dynamic shimming and shim arrays with 32 and 48 channels. Results The eight-channel shim array provided 12% improvement in whole brain static shimming and provided 33% improvement when using slice-based shimming. With this, the eight-channel array performed similar to third-order dynamic shimming (without the need for higher order eddy current compensation). More complex shim arrays with 32 and 48 channels performed better, but require a dedicated RF coil. Discussion The designed eight-channel shim array provides a low-complexity and low-cost approach for improving B0 field shimming on an ultra-high field system. In both static and dynamic shimming, it provides improved B0 homogeneity over standard shimming.