Radiation Dose of Contrast-Enhanced Mammography: A Two-Center Prospective Comparison

This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Green circle
Published version: archiving allowed
Data provided by SHERPA/RoMEO

Abstract

The radiation dose associated with contrast-enhanced mammography (CEM) has been investigated only by single-center studies. In this retrospective study, we aimed to compare the radiation dose between two centers performing CEM within two prospective studies, using the same type of equipment. The CEM mean glandular dose (MGD) was computed for low energy (LE) and high energy (HE) images and their sum was calculated for each view. MGD and related parameters (entrance dose, breast thickness, compression, and density) were compared between the two centers using the Mann–Whitney test. Finally, per-patient MGD was calculated by pooling the two datasets and determining the contribution of LE and HE images. A total of 348 CEM examinations were analyzed (228 from Center 1 and 120 from Center 2). The median total MGD per view was 2.33 mGy (interquartile range 2.19–2.51 mGy) at Center 1 and 2.46 mGy (interquartile range 2.32–2.70 mGy) at Center 2, with a 0.15 mGy median difference (p < 0.001) equal to 6.2%. LE-images contributed between 64% and 77% to the total patient dose in CEM, with the remaining 23–36% being associated with HE images. The mean radiation dose for a two-view bilateral CEM exam was 4.90 mGy, about 30% higher than for digital mammography.