Published in

MDPI, Materials, 11(14), p. 3015, 2021

DOI: 10.3390/ma14113015

Links

Tools

Export citation

Search in Google Scholar

Systematic Review and Meta-Analysis of the Effectiveness of Calcium-Phosphate Coating on the Osseointegration of Titanium Implants

This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Green circle
Published version: archiving allowed
Data provided by SHERPA/RoMEO

Abstract

Ca-P coatings on Ti implants have demonstrated good osseointegration capability due to their similarity to bone mineral matter. Three databases (PubMed, Embase, and Web of Science) were searched electronically in February 2021 for preclinical studies in unmodified experimental animals, with at least four weeks of follow-up, measuring bone-to-implant contact (BIC). Although 107 studies were found in the initial search, only eight experimental preclinical studies were included. Adverse events were selected by two independent investigators. The risk of bias assessment of the selected studies was evaluated using the Cochrane Collaboration Tool. Finally, a meta-analysis of the results found no statistical significance between implants coated with Ca-P and implants with etched conventional surfaces (difference of means, random effects: 5.40; 99% CI: −5.85, 16.65). With the limitations of the present review, Ca-P-coated Ti surfaces have similar osseointegration performance to conventional etched surfaces. Future well-designed studies with large samples are required to confirm our findings.