Dark-field computed tomography reaches the human scale

This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Red circle
Preprint: archiving forbidden
Green circle
Postprint: archiving allowed
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

Significance X-ray computed tomography (CT) is one of the most commonly used diagnostic three-dimensional imaging modalities today. Conventionally, this noninvasive technique generates contrast by measuring the X-ray attenuation properties of different tissues. Considering the wave nature of X-rays, complementary contrast can be achieved by further measuring their small-angle scattering (dark-field) properties. This provides additional valuable diagnostic information on otherwise unresolved tissue microstructure. In our work, we have translated this wave-optical mechanism from the optical bench to a human-sized prototype CT system. This involved the integration of an interferometer into a clinical CT gantry and overcoming several associated challenges regarding vibrations, continuous gantry rotation, and large field of view. This development puts complementary X-ray contrast within reach for real-word medical applications.