Published in

MDPI, Applied Sciences, 4(11), p. 1451, 2021

DOI: 10.3390/app11041451

Links

Tools

Export citation

Search in Google Scholar

UHPLC-MS Method for the Analysis of the Molecular Adjuvant Sulfavant A

This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Green circle
Published version: archiving allowed
Data provided by SHERPA/RoMEO

Abstract

A fast and sensitive method that is based on Ultra High Performance Liquid Chromatography coupled with High Resolution Mass Spectrometry (UHPLC-HRMS) for the measurement of Sulfavant A, a molecular adjuvant with a sulfolipid skeleton, is described. The method has been validated over the linearity range of 2.5–2000 ngmL−1 using a deuterated derivative (d70-Sulfavant A) as internal standard. Chromatographic separation is based on a UHPLC Kinetex® 2.6 µm PS C18 column and a gradient of methanol in 0.32 mM ammonium hydroxide solution buffered at pH 8. The lowest limit of quantification of Sulfavant A was 6.5 ngmL−1. The analytical procedure was tested on an extract of mice lung spiked with 30, 300, and 1500 ng of Sulfavant A. The analysis revealed a precision and accuracy value (as a mean value of all the quality control samples analyzed) of 4.7% and 96% in MeOH and 6.4% and 93.4% in the lung extracts, respectively.