Wiley, Magnetic Resonance in Medicine, 6(67), p. 1506-1514, 2011
DOI: 10.1002/mrm.23136
Full text: Download
Motion-induced artefacts in magnetic resonance spectroscopic imaging are much harder to recognize than in imaging experiments and can therefore lead to erroneous interpretation. A method for prospective motion correction based on an optical tracking system has recently been proposed and has already been successfully applied to single voxel spectroscopy. In this work, the utility of prospective motion correction in combination with retrospective phase correction is evaluated for spectroscopic imaging in the human brain. Retrospective phase correction, based on the interleaved reference scan method, is used to correct for motion-induced frequency shifts and ensure correct phasing of the spectra across the whole spectroscopic imaging slice. It is demonstrated that the presented correction methodology can reduce motion-induced degradation of spectroscopic imaging data.