Dissemin is shutting down on January 1st, 2025

Published in

Sociedade Brasileira de Química, SBQ, Journal of the Brazilian Chemical Society, 2021

DOI: 10.21577/0103-5053.20200235

Links

Tools

Export citation

Search in Google Scholar

Alternative Approaches Applied to Inductively Coupled Plasma Techniques: Multi‑Flow and Two-Flow Calibration

This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Question mark in circle
Preprint: policy unknown
Question mark in circle
Postprint: policy unknown
Question mark in circle
Published version: policy unknown
Data provided by SHERPA/RoMEO

Abstract

Multi-flow calibration (MFC) is based on monitoring the analytical signal from a single calibration standard solution at several different nebulization gas flow rates (Q), which normalizes plasma conditions and minimizes matrix effects. In the present study, MFC was evaluated, for the first time, applied to inductively coupled plasma optical emission spectrometry (ICP OES) and inductively coupled plasma mass spectrometry (ICP-MS) to analyze different and complexmatrix samples. Al, As, Cd, Cr, Cu, Mn, Pb, and Zn were determined by ICP OES, while As, Cd, Co, Cu, Hg, Mo, Mn, Ni, Se, Sb, Pb, and V were determined by ICP-MS. MFC results were compared with those obtained using external standard calibration (EC) for both techniques and in all cases, MFC showed equal or superior accuracy (recoveries between 80-120%) compared to EC, and lower relative standard deviation (RSDs ≤ 10%). Several tests were also performed using only two nebulization gas flows to build linear models for calibration (called two-flow calibration, TFC) and the accurate results (recoveries ranged from 80 to 110% for ICP OES and from 81 to 102% for ICP-MS) suggests that this strategy can be also applied, resulting in a method with high sample throughput.