Published in

BioMed Central, BMC Research Notes, 1(13), 2020

DOI: 10.1186/s13104-020-05355-0

Links

Tools

Export citation

Search in Google Scholar

Direct application of an ECG-based sleep staging algorithm on reflective photoplethysmography data decreases performance

This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Green circle
Published version: archiving allowed
Data provided by SHERPA/RoMEO

Abstract

Abstract Objective The maturation of neural network-based techniques in combination with the availability of large sleep datasets has increased the interest in alternative methods of sleep monitoring. For unobtrusive sleep staging, the most promising algorithms are based on heart rate variability computed from inter-beat intervals (IBIs) derived from ECG-data. The practical application of these algorithms is even more promising when alternative ways of obtaining IBIs, such as wrist-worn photoplethysmography (PPG) can be used. However, studies validating sleep staging algorithms directly on PPG-based data are limited. Results We applied an automatic sleep staging algorithm trained and validated on ECG-data directly on inter-beat intervals derived from a wrist-worn PPG sensor, in 389 polysomnographic recordings of patients with a variety of sleep disorders. While the algorithm reached moderate agreement with gold standard polysomnography, the performance was significantly lower when applied on PPG- versus ECG-derived heart rate variability data (kappa 0.56 versus 0.60, p < 0.001; accuracy 73.0% versus 75.9% p < 0.001). These results show that direct application of an algorithm on a different source of data may negatively affect performance. Algorithms need to be validated using each data source and re-training should be considered whenever possible.