Several retrieval studies have reported on metallic depositions on ceramic femoral heads, but the effect on the wear behavior of artificial hip joints has not been investigated in wear simulator studies. In the present study, retrieved ceramic heads with metallic depositions as third particles were tested against cross-linked ultra-high-molecular-weight polyethylene (UHMWPE) liners in a hip wear simulator. The amount of liner wear and expansion of metallic depositions on the heads were determined before and after wear testing with digital microscopy. The surface roughness of the heads was investigated in areas with and without metallic depositions by laser scanning microscopy. After five million load cycles, a non-significant reduction in the metallic formation on the retrieved heads was found. The metallic areas showed a higher surface roughness compared to unconcerned areas. The liners showed a higher wear rate of 1.57 ± 1.36 mg/million cycles for 28 mm heads and 2.42 ± 0.82 mg/million cycles for 36 mm heads with metallic depositions, in comparison with new ceramic heads with a 28 mm size ((−0.06 ± 0.89) mg/million cycles) and 36 mm size ((2.04 ± 0.46) mg/million cycles). Metallic transfer on ceramic heads can lead to an increased surface roughness and higher wear rates at the UHMWPE liners. Therefore, metallic contact of the ceramic femoral head should be avoided.