Paclitaxel-Coated Balloon vs Uncoated Balloon Angioplasty for Treatment of In-Stent Restenosis in the Superficial Femoral and Popliteal Arteries: The COPA CABANA Trial

This paper was not found in any repository, but could be made available legally by the author.
This paper was not found in any repository, but could be made available legally by the author.

Full text: Unavailable

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

Purpose: To investigate the efficacy and sustainability of drug-coated balloon (DCB) treatment of femoropopliteal in-stent restenosis (ISR). Materials and Methods: An investigator-initiated, prospective, multicenter, 1:1 randomized study enrolled 88 patients for treatment of ISR with DCB (n=47; mean age 68.3±9.6 years; 26 men) or uncoated balloon (n=41; mean age 67.6±10.2 years; 26 men) angioplasty ( ClinicalTrials.gov identifier NCT01594684). Additionally, the protocol provided for an observational arm composed of patients from either randomized arm who experienced recurrent ISR ≥30 days after the index treatment. Redo treatment consisted of 2 DCBs sequentially inflated at the same location (double dose therapy). The majority of patients (66, 78%) had Rutherford category 3 ischemia. The mean lesion length was 140 mm; a third (27, 31%) were total occlusions. The primary endpoint was angiographic late lumen loss (LLL) at 6 months evaluated by an independent core laboratory. Results: Twenty-two patients (7 DCB +15 uncoated) were treated for recurrence with fully overlapping double DCB angioplasty. Six-month LLL was lower after DCB (0.34±1.12 mm) treatment than after angioplasty with an uncoated balloon (1.58±1.10 mm, p<0.001). At the 12-month follow-up, target lesion revascularization (TLR) was performed in 18 (49%) of 37 patients in the uncoated group, 6 (14%) of 43 patients in the single-dose DCB group (p=0.001), and no patients from the recurrent ISR group. At ~2 years after treatment, a remarkable number (14/27, 52%) of TLRs were recorded in the single-dose DCB group. Conclusion: Treatment with DCBs resulted in significantly less 6-month LLL and fewer TLRs up to 24 months than treatment with uncoated balloons. The double dose for treating recurrent ISR did not cause recognizable adverse events or require TLR up to 24 months.