Published in

Millimeter, Submillimeter, and Far-Infrared Detectors and Instrumentation for Astronomy V

DOI: 10.1117/12.857871

Links

Tools

Export citation

Search in Google Scholar

The Keck Array: a pulse tube cooled CMB polarimeter

This paper is available in a repository.
This paper is available in a repository.

Full text: Download

Red circle
Preprint: archiving forbidden
Green circle
Postprint: archiving allowed
Green circle
Published version: archiving allowed
Data provided by SHERPA/RoMEO

Abstract

The Keck Array is a cosmic microwave background (CMB) polarimeter that will begin observing from the South Pole in late 2010. The initial deployment will consist of three telescopes similar to BICEP2 housed in ultra-compact, pulse tube cooled cryostats. Two more receivers will be added the following year. In these proceedings we report on the design and performance of the Keck cryostat. We also report some initial results on the performance of antenna-coupled TES detectors operating in the presence of a pulse tube. We find that the performance of the detectors is not seriously impacted by the replacement of BICEP2's liquid helium cryostat with a pulse tube cooled cryostat. ; Comment: 8 pages, 6 figures; SPIE proceedings for Millimeter, Submillimeter and Far-Infrared Detectors and Instrumentation for Astronomy V (Conference 7741, June 2010, San Diego, CA, USA)