Direct Comparison of the Thickness of the Parietal Peritoneum Using Peritoneal Biopsy and Ultrasonography of the Abdominal Wall in Patients Treated with Peritoneal Dialysis

This paper was not found in any repository; the policy of its publisher is unknown or unclear.
This paper was not found in any repository; the policy of its publisher is unknown or unclear.

Full text: Unavailable

Question mark in circle
Preprint: policy unknown
Question mark in circle
Postprint: policy unknown
Question mark in circle
Published version: policy unknown
Data provided by SHERPA/RoMEO

Abstract

BackgroundLong-term treatment with peritoneal dialysis (PD) results in peritoneal fibrosis. Peritoneal biopsies have been used to determine the severity of fibrosis. Ultrasonography (US) of the abdominal wall has been used to measure peritoneal thickness non-invasively. However, direct comparison of both methods in the same patient has never been done. Furthermore, the validity of US to measure peritoneal thickness has not been investigated.MethodsWe performed 3 studies: 1) a human biopsy study to compare US measurement of peritoneal thickness with histological examination; 2) a human cadaver study to investigate the effect of removing the peritoneum on US results; and 3) a phantom study in which we used US to measure the thickness of membrane-like structures with a known thickness to investigate the influence of different US settings.ResultsThe median thickness in biopsies of the peritoneum was 113 μm (interquartile range [IQR] 72 –129 μm), while this was 370 μm (IQR 324 – 458 μm) when measured by US ( p < 0.0001). The mean difference between the 2 measures was -257 μm (limits of agreement -4.6 and -511 μm). In the cadaver study, removal of the peritoneum did not have an effect on the presence or thickness of the hyperechoic line reported to represent the peritoneum. In the phantom study, results were highly dependent on frequency of the transducer, scan depth, and gain settings.ConclusionsUltrasonography results differ markedly from histological measurement using peritoneal biopsies. However, the hyperechoic line generated by US represents the interface between 2 neighboring tissues and not a separate morphological structure. Moreover, its thickness is greatly influenced by user-defined US settings.