Fractional flow reserve derived from microcatheters versus standard pressure wires: a stenosis-level meta-analysis

This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Green circle
Published version: archiving allowed
Data provided by SHERPA/RoMEO

Abstract

AimsTo determine the agreement between sensor-tipped microcatheter (MC) and pressure wire (PW)-derived fractional flow reserve (FFR).Methods and resultsStudies comparing FFR obtained from MC (FFRMC, Navvus Microcatheter System, ACIST Medical Systems, Eden Prairie, Minnesota, USA) versus standard PW (FFRPW) were identified, and a meta-analysis of numerical and categorical agreement was performed. The relative levels of drift and device failure of MC and PW systems from each study were assessed. Six studies with 440 lesions (413 patients) were included. The mean overall bias between FFRMC and FFRPW was −0.029 (FFRMC lower). Bias and variance were greater for lesions with lower FFRPW (p<0.001). Using a cut-off of 0.80, 18 % of lesions were reclassified by FFRMC versus FFRPW (with 15 % being false positives). The difference in reported drift between FFRPW and FFRMC was small. Device failure was more common with MC than PW (7.1% vs 2%).ConclusionFFRMC systematically overestimates lesion severity, with increased bias in more severe lesions. Using FFRMC changes revascularisation guidance in approximately one out of every five cases. PW drift was similar between systems. Device failure was higher with MC.