Dissemin is shutting down on January 1st, 2025

Published in

MDPI, Sensors, 5(13), p. 5841-5856, 2013

DOI: 10.3390/s130505841

Links

Tools

Export citation

Search in Google Scholar

A Ferrocene-Porphyrin Ligand for Multi-Transduction Chemical Sensor Development

This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Green circle
Published version: archiving allowed
Data provided by SHERPA/RoMEO

Abstract

5,10,15,20-Tetraferrocenyl porphyrin, H2TFcP, a simple example of a donor-acceptor system, was tested as ligand for the development of a novel multi-transduction chemical sensors aimed at the determination of transition metal ions. The fluorescence energy transfer between ferrocene donor and porphyrin acceptor sub-units was considered. The simultaneously measured optical and potentiometric responses of solvent polymeric membranes based on H2TFcP permitted the detection of lead ions in sample solutions, in the concentration range from 2.7 × 10−7 to 3.0 × 10−3 M. The detection limit of lead determination was 0.27 μM, low enough to perform the direct analysis of Pb2+ in natural waters.