Published in

American Institute of Physics, Review of Scientific Instruments, 1(88), p. 013901

DOI: 10.1063/1.4973419

Links

Tools

Export citation

Search in Google Scholar

Simple quadratic magneto-optic Kerr effect measurement system using permanent magnets

Journal article published in 2017 by A. V. Pradeep, Sayak Ghosh ORCID, P. S. Anil Kumar
This paper was not found in any repository, but could be made available legally by the author.
This paper was not found in any repository, but could be made available legally by the author.

Full text: Unavailable

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Orange circle
Published version: archiving restricted
Data provided by SHERPA/RoMEO

Abstract

In recent times, quadratic magneto-optic Kerr effect (QMOKE) is emerging as an important experimental tool to investigate higher-order spin-orbit interactions in magnetic thin films and heterostructures. We have designed and constructed a simple, cost-effective QMOKE measurement system using permanent magnets. The permanent magnets are mounted on the inner surface of a cylindrical ferromagnetic yoke which can be rotated about its axis. Our system is sensitive to both the quadratic and linear MOKE signals. We use rotating field method to extract the QMOKE components in saturation. This system is capable of extracting the QMOKE signal from single crystals and thin film samples. Here we present the construction and working of the QMOKE measurement system using permanent magnets and report, for the first time, the QMOKE signal from Fe3O4 single crystal.