Published in

Lavoisier, Journal of Water Science / Revue des Sciences de l'Eau, 3(11), p. 409-428, 2005

DOI: 10.7202/705314ar

Links

Tools

Export citation

Search in Google Scholar

Étude du colmatage de membranes d'osmose inverse utilisées pour la préparation des eaux de dialyse rénale

Journal article published in 1998 by E. Roth, B. Fabre, A. Accary, B. Faller
This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

L'osmose inverse est l'élément clef de la production d'eau ultrapure servant dans la préparation du liquide de dialyse rénale. Le vieillissement des membranes d'osmose inverse se traduit par une dérive croissante de la conductivité en sortie de membrane et par une augmentation de la teneur en bactéries dans le perméat délivré en sortie d'osmose inverse, entraînant un risque de contamination pour le patient.Les suivis en conductivité et numérations bactériennes dans la chaîne de production d'eau osmosée de l'Hôpital de Colmar (France) montre un fort dysfonctionnement au niveau des prétraitements et notamment du charbon actif, siège d'une importante prolifération bactérienne (2.3 103 colonies pour 100 ml en amont du charbon actif , 1.1 105 pour 100 ml en aval). Une étude de corrélations entre entrée et sortie d'osmoseurs montre que la forte teneur en germes en sortie du charbon actif est directement responsable de la présence des bactéries dans le perméat.Des tests hydrauliques effectués sur les membranes après quatre années d'utilisation intensive (6500 heures par an), permettent d'identifier les mécanismes prépondérants de vieillissement et de colmatage des membranes utilisées à l'hôpital de Colmar (compaction, formation d'un dépôt et altération de structure).Afin d'identifier les substances colmatantes non retenues par le prétraitement, le film colmatant est analysé par analyse élémentaire, spectrophotométrie infrarouge, diffraction X et zêtamétrie, ce qui nous a permis de mettre en évidence la présence notamment d'argiles, de silice et de substances humiques. ; The use of reverse osmosis to produce ultrapure water for kidney patients is very prevalent at the present time. About 20000 patients are treated every day in France and the cost of treatment is estimated at 10 thousand million French francs. The patients are provided with a permeate which in principle contains neither ions nor bacteria, which can cause disease and irreversible alteration of the upper-arm articulations. After several months of utilization, the presence of a foulant deposit and biofilm proliferation is often noted on the thin layered active surface of the polyamide membranes. In many cases, the membrane may become irreversibly fouled and require replacement.The aim of this work was to determine the causes of membrane fouling at a reverse osmosis plant in "Hôpital Pasteur" (Colmar, France). First we analyzed the ion-rejection capacity of membranes and the evolution of conductivity over 200 weeks. We observed an increasing drift in conductivity, which could not be halted by cleaning and regeneration procedures. In the second phase of the study, we carried out microbiological studies at many sampling points of the ultrapure water plant in "Hôpital Pasteur". We noted a very important proliferation of bacteria downstream of the pretreatments. The activated charcoal unit is responsible for this problem (2.3 x 103 UFC / 100 mL upstream from the activated charcoal bed, 1.1 x 105 UFC / 100 mL downstream). We noted an important dysfunction in the pretreatment, essentially attributable to the activated charcoal (bacteria) and the resins (silica).We performed tests with x-ray diffraction as a rapid method to determine the crystalline forms present. In the foulant deposits, we found quartz and clay. Infrared spectroscopy is useful in determining the type of organic constituents; in our deposits we noted the presence of humic materials. The analysis of those deposits gave essentially carbon and silicon. The foulant deposits were of organic and crystalline origins.