Published in

Faculdade de Letras, Brazilian Journal of Pharmaceutical Sciences, 3(46), p. 551-554, 2010

DOI: 10.1590/s1984-82502010000300019

Links

Tools

Export citation

Search in Google Scholar

Influence of radiation on endotoxin test using the PTS TM for 18-FDG radiopharmaceutical

Journal article published in 2010 by Ralph Santos-Oliveira ORCID
This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Green circle
Published version: archiving allowed
Data provided by SHERPA/RoMEO

Abstract

F-18 FDG (2-[18-F] fluoro-2-deoxy-D-glucose) is the most frequently used radiopharmaceutical for PET and PET CT imaging exams. The FDA recently approved the use of the PTS TM (Portable Test System) as an alternative to the standard test proposed by the United States Pharmacopeia using the LAL (Limulus Amebocyte Lysates), that takes longer to perform (about 1h) than the PTS TM (15 min). Recent studies have demonstrated that radiation could interfere with the PTS TM test. In order to study the effects of radiation on the PTS TM test and/or equipment, 27 batches of F-18 FDG produced in the Nuclear Engineering Institute were analyzed. The results showed that no direct correlation with radiation was found in any of the cases.