Published in

Medknow Publications, Annals of Cardiac Anaesthesia, 1(16), p. 54, 2013

DOI: 10.4103/0971-9784.105372

Links

Tools

Export citation

Search in Google Scholar

Magnets and implantable cardioverter defibrillators: What′s the problem?

Journal article published in 2013 by Yiliam F. Rodriguez Blanco, Fouad Souki, Evelyn Tamayo, Keith Candiotti
This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Green circle
Published version: archiving allowed
Data provided by SHERPA/RoMEO

Abstract

A growing number of surgical patients present to the operating room with implantable cardioverter defibrillators (ICD). Peri-operative care of these patients dictates that ICD function be suspended for many surgical procedures to avoid inappropriate, and possibly harmful, ICD therapy triggered by electromagnetic interference (EMI). An alternative to reprogramming the ICD is the use of a magnet to temporarily suspend its function. However, this approach is not without complications. We report a case where magnet use failed to inhibit ICD sensing of EMI, and a shock was delivered to the patient. Measures to decrease EMI, controversies regarding magnet use, and expert recommendations are discussed.