Published in

Wiley, Journal of Microscopy, 2(205), p. 118-124

DOI: 10.1046/j.0022-2720.2001.00981.x

Links

Tools

Export citation

Search in Google Scholar

A technique for the examination of polar ice using the scanning electron microscope

Journal article published in 2002 by P. R. F. Barnes ORCID, R. Mulvaney ORCID, E. W. Wolff, K. Robinson
This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Green circle
Preprint: archiving allowed
Orange circle
Postprint: archiving restricted
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

The microstructure and location of impurities in polar ice are of great relevance to ice core studies. We describe a reliable method to examine ice in the scanning electron microscope (SEM). Specimens were cut in a cold room and could have their surfaces altered by sublimation either before (pre-etching) or after (etching) introduction to the cryo-chamber of the SEM. Pre-etching was used to smooth surfaces, whilst etching stripped away layers from the specimen surface, aiding the location of particles in situ, and allowing embedded structures to be revealed. X-ray analysis was used to determine the composition of localized impurities, which in some cases had been concentrated on the surface by etching. Examining uncoated surfaces was found to be advantageous and did not detract from qualitative X-ray analysis. Imaging uncoated was performed at low accelerating voltages and probe currents to avoid problems of surface charging.