Fast volumetric fluorescence imaging with multimode fibers

This paper was not found in any repository, but could be made available legally by the author.
This paper was not found in any repository, but could be made available legally by the author.

Full text: Unavailable

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

In this Letter, we propose a compact multimode fiber endoscope which employs wavefront shaping with a digital micromirror device (DMD). An automated single calibration step allows us to correct for optical misalignment, and the method achieves accurate focusing at various depths in the sample through rapid switching of holographic patterns by the DMD. The speed of calibration is one or two orders of magnitude faster than existing methods. The method, single calibration multimode fiber imaging (SCMFI), is compared with existing methods, and its performance is validated. We show a near diffraction limited focusing capability at imaging depths up to 110 µm with near constant lateral resolutions of 1.4 µm. Finally, we demonstrate the method for the imaging of small fluorescent beads embedded in a 3D matrix. The results indicate excellent power penetration and focusing performance. Combined with the high speed of SCMFI, this paves the way for volumetric tissue endoscopy at depth.