Reflections on the Projection of Ions in Atom Probe Tomography

Journal article published in 2017 by Frédéric De Geuser, Baptiste Gault ORCID
This paper is available in a repository.
This paper is available in a repository.

Full text: Download

Green circle
Preprint: archiving allowed
Red circle
Postprint: archiving forbidden
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

AbstractThere are two main projections used to transform, and reconstruct, field ion micrographs or atom probe tomography data into atomic coordinates at the specimen surface and, subsequently, in three dimensions. In this article, we present a perspective on the strength of the azimuthal equidistant projection in comparison with the more widely used and well-established point projection (or pseudo-stereographic projection), which underpins data reconstruction in most software packages currently in use across the community. After an overview of the reconstruction methodology, we demonstrate that the azimuthal equidistant is more robust with regards to errors on the parameters used to perform the reconstruction and is therefore more likely to yield more accurate tomographic reconstructions.