Links

Tools

Export citation

Search in Google Scholar

Measurements of Cardiac Output during Constant Exercises: Comparison of Two Non-Invasive Techniques

Journal article published in 2004 by N. Tordi, L. Mourot, B. Matusheski, R. L. Hughson ORCID
This paper is available in a repository.
This paper is available in a repository.

Full text: Download

Question mark in circle
Preprint: policy unknown
Question mark in circle
Postprint: policy unknown
Question mark in circle
Published version: policy unknown

Abstract

We compared cardiac output (CO) determined simultaneously by electrical impedance cardiography method (CO (ICG)) and by the CO (2) rebreathing technique (CO (2REB)) during three different steady-state exercises (target heart rate of 120, 140, and 160 min (-1)) in 8 healthy fit young men. The mean difference correlation coefficient obtained between the values of CO (ICG) and CO (2REB) was 0.85 and the mean difference (CO (ICG)-CO (2REB)) was 0.06 l/min (0.12 %). At 120 min (-1), CO (ICG) was lower than CO (2REB) but the tendency was reversed at 140 and 160 min (-1) where CO (ICG) was higher than CO (2REB). This evolution may be explained by the difficulty of using CO (2) rebreathing technique at the highest steady-state exercises and by the progressive acidemia due to exercise. The present results suggest that electrical impedance cardiography method provides acceptable evaluation of CO and may favourably replace the CO (2) rebreathing technique during mild (or moderate) to high steady-state exercises.