Published in

Journal of Novel Physiotherapies, 01(03)

DOI: 10.4172/2165-7025.1000126

Links

Tools

Export citation

Search in Google Scholar

A Comparison of VO2 Measurement Obtained by a Physiological Monitoring Device and the Cosmed Quark CPET

Journal article published in 2013 by Kelly A. Brooks, Jeremy G. Carter, James J. Dawes
This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

White circle
Preprint: policy unclear
Green circle
Postprint: archiving allowed
Green circle
Published version: archiving allowed
Data provided by SHERPA/RoMEO

Abstract

The purpose of this study was to compare the results of measurements of VO2 as obtained by the Zephyr Bioharness (ZB) and the Cosmed Quark CPET Metabolic cart (CM). Both ZB and CM have been proven reliable by previous research. The ZB is portable and offers many practical advantages over CM, which requires greater expertise and expense, and tests must be completed in a laboratory setting. Aerobic capacity was assessed in a sample of college students (n=33). Subjects performed a maximal test using a treadmill protocol. Mean Ventilation (VE), mean heart rates (HR), respiratory exchange ratios (RER), and aerobic capacity (VO2) were similar using both the ZB and the CM, with no significant differences observed. Because data varied very little, both methods are comparable, which indicates ZB is a practical solution for monitoring physiological variables outside of a laboratory setting.